Canada
270 Sherman Avenue N.
Hamilton, ON L8L 6N5
Phone: (905) 545 2003• Fax: (905) 545 4248
United States
1925 Pine Avenue
Niagara Falls, NY 14301
Phone: (716) 285-0418 • Fax: (716) 282-2937
 
Email: sales@nova-gas.com or websales@nova-gas.com
All Categories > Multigas Analyzers > Item # 352 Series  


Item # 352 Series, Portable Flue Gas Analyzer for Oxygen, Carbon Monoxide and Combustibles


The Nova 352 Series Portable Flue Gas Analyzer has been designed for accuracy, reliability, ease of use and ease of service. It uses customer replaceable sensors which respond quickly to the oxygen (O2),  carbon monoxide (CO) and combustibles present in the flue gas sample. The sensor life expectancy is between 3 and 4 years.

In operation, a built in sample pump draws in the flue gas sample through the stainless steel probe, 12 ft. (4 m) sample hose, condensate removal filter, secondary filter, flow meter, then on to the O2, CO and combustibles sensors. The detected gases are displayed on LCD digital meters which have a switchable backlight for use in dark areas. A built in air makeup system ensures that the combustibles detector will always have sufficient O2 for proper operation regardless of sample O2 content.

A rechargeable battery provides enough power for about 6 hours of continuous operation and the analyzer can be used while it is being recharged. A red LED tells when to recharge and a green LED verifies that it is receiving recharging power. The recharger is included.

The Nova 352T indicates stack temperature for doing fuel efficiency calculations. The temperature sensor is built into the sampling probe. Efficiency charts for each fuel are provided




 Specifications 

Method of Detection

Catalytic oxidation detector for combustibles; expected life is 3-4 years for each
Customer replaceable electrochemical O2 and CO sensors

Ranges Available

0-1800°F or 0-1000 °C stach temperature (Model 352T)
0-2000PPM and 4.00% CO (switch selectable)

0-25.0% O2

0-5.0% or 0-10.0% combustibles

Resolution

0.1% O2, CO and combustibles

Accuracy & Repeatability

±1% full scale, based on 20.9% O2; ± 2% of full scale combustibles

Drift

<2% full scale per 8 hours of continuous operation

Response Time (T-90)

5-8 seconds for O2; 20-30 seconds for combustibles

Ambient Temperature Range

0 to 40.5 ºC
32 to 105
ºF

Linearity

±1% full scale O2 and CO; ±2% of full scale combustibles

Size & Weight

Suitcase (K): 14" L x 10 1/2" W x 6" H @ 12 lbs (36 x 27 x 15 cm @ 5.5 kg)
Weather Proof (WP) Case: 10" L x 7 1/2" H x 8" W @ 8 lbs (25 x 19 x 20 cm @ 3.6 kg)

Power

AC/DC Operation, 115 VAC, 60 Hz for recharging (Other voltages available)

Output Options

4-20 mA or 0-1 VDC

Unique Applications

The 352 Series should not be used for detecting these gases in ambient atmospheres for personnel safety purposes.

All Nova analyzers are built using proven technologies and techniques. If this product does not suit your application, please contact Nova. In many cases, we are able to build an analyzer specific to your needs.

Primary Gas to be Measured

Flue Gas or Emissions

Industry Type

Steel Making



Print Back