Canada
270 Sherman Avenue N.
Hamilton, ON L8L 6N5
Phone: (905) 545 2003• Fax: (905) 545 4248
United States
1925 Pine Avenue
Niagara Falls, NY 14301
Phone: (716) 285-0418 • Fax: (716) 282-2937
 
Email: sales@nova-gas.com or websales@nova-gas.com
All Categories > Multigas Analyzers > Item # 3255 Series  


Item # 3255 Series, Portable Analyzer for PPM Oxygen and Dew Point


The Model 3255 Portable Analyzer has been designed for the accurate measurement of both PPM levels of O2 and low-level dew points in furnace atmosphere gas or in other dry process gases.  A customer-replaceable electrochemical sensor senses the O2. This sensor produces a small output current, which is directly proportinal to the PPM level of O2 present in the sample.  The dew point is sensed by an aluminum oxide sensor that baries its electrical capacitance with moisture.  The O2  and dew point signals are then amplified and displayed on the front panel meters.

A unique feature of this analyzer is the built-in micro-blocking valves that trap the last gas sample drawn into the sensors.  This ensures a rapid start and minimum purge-down time when the analyzer is used again.  The optional hydrogen detector is a thermal conductivity (T/C) Cell which responds rapidly to the hydrogen content of the sample gas.  It has an inifinite life.





 Specifications 

Method of Detection

Customer replaceable O2 sensor, AlO2 Dew Point Sensor, T/C Cell for H2

Ranges Available

0-10 PPM to 0-19, 999 PPM O2, -112° to + 68°F (-80° to +20°C)

Resolution

±1PPM O2, ±1°C or °F Dew Point, 0.1% H2

Accuracy & Repeatability

±2% of reading

Drift

Less than 1% of full scale per 8 hours

Response Time (T-90)

O2 sensor may require an initial

Ambient Temperature Range

0-50 ºC
32 to 122
ºF

Linearity

±1% of full scale

Size & Weight

BT Style: 7 1/2"W x 9 1/2" H x 10"D @ 10 lbs (19 x 24 x 25 cm @ 4.5 kg)
K Style: 14"L x 6" H x 10 1/2"D @ 12 lbs (35.5 x 15.2 x 26.6 cm @ 5.5 kg)

WP Style: 10"L x 7 1/2" H x 6 1/2"D @ 8 lbs (25.5 x 19 x 16.5 cm @ 3.6 kg)

Power

AC/DC Operation, 115VAC 60Hz for recharging. (220VAC 50Hz available)

Output Options

4-20 mA or 0-1 VDC

Alarms

Optional

Unique Applications

All Nova analyzers are built using proven technologies and techniques. If this product does not suit your application, please contact Nova. In many cases, we are able to build an analyzer specific to your needs.

Industry Type

Flue Gas or Emissions



Print Back