Method of Detection |
N/A Microprocessor based infrared detector for CO2 |
Ranges Available |
N/A 0-3000, 0-5000 or 0-10,000 PPM; 0-2%, 0-5%, 0-10%, 0-20%, 0-50% or 0-100% CO2. Other ranges available. |
Resolution |
N/A 0.1% CO2 on % versions; 10 PPM CO2 on PPM versions |
Accuracy & Repeatability |
N/A Within ±2% full scale |
Drift |
N/A Within 1% full scale per 8 hours of continuous operation |
Response Time (T-90) |
N/A 30 - 40 seconds |
Ambient Temperature Range |
N/A 4 to 49 ºC40 to 120 ºF |
Linearity |
N/A Better than 1.0% of full scale |
Size & Weight |
N/A approx. 26 x 18 x 13 cm @ 3.4 kg (10" x 7" x 5" @ 7.5lbs) |
Power |
N/A 115 VAC 60 Hz for recharging (220VAC / 50Hz also available) |
Output Options |
N/A 4-20mA |
Hazardous Area Rating Options |
N/A General Purpose, Non-Hazardous |
N/A For checking carbon dioxide (CO2) content in mushroom growing rooms, incubators, natural gas, controlled atmosphere storage rooms, welding gas mixtures, and other process applications. |
N/A
|
N/A
|