Items |
Model 380K Portable Tri Gas Analyzer for Hydrogen Purging, Suitcase (K) Enclosure |
Model 381K Portable Tri-Gas Analyzer for H2 / N2 / Air Purging |
Model 382K Portable Tri Gas Analyzer for Hydrogen Purging with separate O2 channel |
|||
Method of Detection | N/A Temperature compensated thermal conductivity (T/C) cell | N/A Thermal conductivity (T/C) cell, electrochemical O2 sensor | N/A Thermal-conductivity (T/C) cell for H2; Electrochemical sensor for O2 | |||
Ranges Available | N/A Range 1 : 0-100% H2 in Air Range 2 : 0-100% H2 in CO2 Range 3 : 0-100% Air in CO2 |
N/A
Range 1: 0-100% H2 in Air Range 2: 0-100% H2 in N2 Range 3: 0-100% Air in N2 |
N/A
Range 1: 0-100% H2 in Air Range 2: 0-100% H2 in CO2 Range 3: 0-100% Air in CO2 Separate Range: 0-25.0 % O2 |
|||
Resolution | N/A 0.1% of gas measured | |||||
Accuracy & Repeatability | N/A Range 1: ± 1% of Full Scale; Range 2: ± 2% of F.S.; Range 3: ± 2% of F.S. | N/A ±1% F.S except in 85-100% H2 in Air range, which is within 0.5% absolute H2 | N/A Range 1: ± 1% of Full Scale; Range 2 & 3: ± 2% of F.S.; O2 Range: ± 1% of F.S. | |||
Drift | N/A H2 in Air range is ± 0.4% per week H2 in CO2 or Air in CO2 2% F.S. per week maximum drift | N/A H2 in N2 or Air in N2, ± 2% F.S. per week maximum drift; 0-100% H2 in Air range is ± 0.4% per week maximum | N/A H2 in CO2, Air in CO2, & O2 ± 2% F.S. per week maximum drift; 0-100% H2 in Air range is ± 0.4% per week maximum | |||
Response Time (T-90) | N/A 10-15 seconds to 90% step change - not including sample transport time | |||||
Ambient Temperature Range | N/A 0 to 50 ºC32 to 120 ºF | |||||
Linearity | N/A ±1.0% of F.S. on H2 in Air range. ±2% of F.S. in H2 or Air in CO2 ranges | N/A ± 1.0% of F.S. on H2 in Air range. ± 2% of F.S. in H2 or Air in N2 ranges | N/A ± 1.0% of F.S. on H2 in Air & O2 range. ± 2% of F.S. in H2 & Air in CO2 ranges | |||
Size & Weight | N/A 9 1/2" L x 7" W x 6 1/2" H @ 12 lbs (24 x 17 x 18 cm @ 3.6 kg) | |||||
Power | N/A Battery operation. 115 VAC, 60 Hz for recharging (220 VAC, 50 Hz available) | |||||
Output Options | N/A 4-20mA | |||||
Options | N/A Recorder output - 4-20mA Suitcase (K) style cabinet available | N/A Recorder output - 4-20 mA Suitcase (K) style cabinet available | N/A Recorder output - 4-20mA Suitcase (K) style cabinet available | |||
Analyzer Type | N/A Tri-Gas Analyzer | |||||
Hazardous Area Rating Options | N/A General Purpose, Non-Hazardous | |||||
|