New Call-to-action
The Nova 311 Series Portable SO2 Analyzer utilizes a reliable, customer-replaceable electrochemical sensor which responds quickly to the SO2 present in the flue gas sample. Its life expectancy is between 2 and 3 years. The pre-dried gas sample drawn into the analyzer by the sample pump is filtered, and then indicated on the flow meter. The cleaned sample then passes through the sensor chamber where the SO2 is detected and displayed on a digital readout.

A rechargeable battery powers the pump and electronics for about 8 hours of operation. The analyzer can still be operated while it is being charged. Recharger is included.
Unit of Measure

Specifications

Method of Detection

N/A Customer replaceable electrochemical SO2 sensor

Ranges Available

N/A Any range up to 0-2000 PPM SO2

Resolution

N/A ±10 PPM

Accuracy & Repeatability

N/A ±1% of full scale, based on a 0-2000 PPM range

Drift

N/A ±1% full scale per month

Response Time (T-90)

N/A 20-30 seconds for 90% of step change

Ambient Temperature Range

N/A 0 to 40 ºC32 to 105 ºF

Linearity

N/A ±1% of full scale

Size & Weight

N/A approx. 26 x 18 x 13 cm @ 3.4 kg (10" x 7" x 5" @ 7.5lbs)

Power

N/A 12V battery operation. 115VAC 60Hz for recharging (220VAC 50Hz available)

Output Options

N/A 4-20mA

Hazardous Area Rating Options

N/A General Purpose, Non-Hazardous

Applications

N/A

        Analysis of sulfur dioxide (SO2) by electrochemical cell in pre-cleaned and dried flue gas and process streams.

Features

N/A

  • Rugged design that is easy to operate and maintain
  • Fast warm-up and response
  • Disposable, long life electrochemical SO2 sensor
  • Digital readout meter
  • Modular layout that is easy to maintain
  • Rechargeable battery operation
  • Any range 0-200 up to 0-2000 PPM SO2
  • Built in sample pump, filter, flow meter, and active condensate removal
  • Rapid reading recovery after SO2 'overdose'
  • Stainless steel probe with sample hose
  • Suitcase cabinet (K) weatherproof when closed

Options

N/A

  • Recorder output - 4-20mA
  • Suitcase-style (K) enclosure
  • Sample pre-cooler
  • On-board data-logging to USB stick

Calibration

N/A

  • Air for zero.
  • Analyzed calibration gas mixture of PPM SO2 in nitrogen for span.

Unique Applications

N/A

    All Nova analyzers are built using proven technologies and techniques. If this product does not suit your application, please contact Nova. In many cases, we are able to build an analyzer specific to your needs.

View Our
Product Catalog
Have a question?
Contact Us
Ready to get started?
Request Quote